Volltext-Downloads (blau) und Frontdoor-Views (grau)

Bitte verwenden Sie diesen Link, wenn Sie dieses Dokument zitieren oder verlinken wollen: https://nbn-resolving.org/urn:nbn:de:gbv:9-opus-112002

Prospective Multicenter Trial of Cervical Arthroplasty with the ROTAIO® Cervical Disc Prosthesis

  • Study design Clinical observational study. Objective The ROTAIO® cervical disc prosthesis is a novel unconstrained implant with a variable center of rotation aiming at physiological motion. The objective of this multicenter prospective trial was to evaluate clinical outcome and complications within 2 years. Material and Methods 120 patients (72 females and 48 males with median age of 43.0 years [23-60 yrs] underwent ACDA (ROTAIO®, SIGNUS Medical, Alzenau, Germany) and were prospectively followed for 24 months. Preoperative complaints were mainly associated with radiculopathy (n = 104) or myelopathy (n=16). There were 108 monosegmental and 12 bisegmental procedures including 6 hybrid constructs. Clinical outcome was evaluated at 3, 12 and 24 months in 100%, 96% and 77% of the cohort by VAS, NDI, WL-26, Patient`s Satisfaction Index (PSI), SF-36, Nurick Score, mJOA, Composite Success Rate, complications, patient`s overall satisfaction and analgesics use. Results Highly significant clinical improvements were observed according to NDI and VAS (P < .0001 (arm); P < .001 (neck); P = .002 (head)) at all time points. Analgetic use could be reduced in 87.1 to 95.2%. Doctor`s visits have been reduced in 93.8% after 24 months. Patient`s overall satisfaction was high with 78.4 to 83.5% of patients. The composite success rate was 77.5% after 12 months and 76.9% after 24 months. There were no major complications in this series. Slight subsidence of the prosthesis was observed in 2 patients and 3 patients demonstrated fusion after 24 months. 2 patients developed symptomatic foraminal stenosis, so that implant removal and fusion was performed resulting in a revision rate of 1.7% in 2 years. Conclusion The ROTAIO® cervical disc prosthesis is a safe and efficient treatment option for symptomatic degenerative disc disease demonstrating highly significant clinical improvement and high patient`s overall satisfaction with very low revision rates at 2 years.

Download full text files

Export metadata

Additional Services

Search Google Scholar
Metadaten
Author: Steffen FleckORCiD, Anna Lang, Jens Lehmberg, Julia F. Landscheidt, Ruediger Gerlach, Julian Rathert, Christian Ulrich, Ralph T. SchärORCiD, Sebastian Hartmann, Jan-Uwe Mueller, Claudius Thome
URN:urn:nbn:de:gbv:9-opus-112002
DOI:https://doi.org/10.1177/21925682221109563
ISSN:2192-5690
Parent Title (English):Global Spine Journal
Publisher:SAGE Publications
Place of publication:Thousand Oaks, CA
Document Type:Article
Language:English
Date of Publication (online):2022/08/05
Date of first Publication:2024/03/01
Release Date:2024/06/28
Tag:cervical arthroplasty; cervical disc prosthesis; complications; djacent segment disease; outcome
Volume:14
Issue:2
First Page:429
Last Page:437
Faculties:Universitätsmedizin / Klinik und Poliklinik für Neurochirurgie
Collections:weitere DFG-förderfähige Artikel
Licence (German):License LogoCreative Commons - Namensnennung 4.0 International